OMEGA Model CYD201/CYD208 User’s Manual
Remote Operation 4-7
WS Sample Sensor Reading and Alarm Status.
Input:
WS
Returned:
For a Model CYD201, returns:
[sign],[sensor reading],[units],[alarm status] (CR)(LF)
For a Model CYD208, returns:
[current channel],[sign],[sensor reading],[units], [alarm
status](CR)(LF)
Remarks:
Returns the sample sensor reading and alarm status, where
A
= active and
I
= inactive.
WY Scan Status
(Model CYD208 Only)
.
Input:
WY
Returned: [scan status],[current channel number],[channel which
caused the alarm state], [channel 1 dwell time],[channel 2
dwell time],...[channel 8 dwell time](CR)(LF)
Remarks:
Returns instrument scan status (scanning or holding),
channel dwell information, and scan position. Scan Status
Parameter:
H
= holding,
S
= scanning. Channel Which
Caused the Alarm State Parameter = 1 through 8 or "-" if
alarm is inactive. Channel dwell times are in seconds.
Sending this command with a Model CYD201 returns
N
to
signify no scanner.